Surface Resistance Meter



               001-9801


               Description
               Measuring resistance: 10  -10 Ω / Surface resistivity: 10  -10 Ω/m 2
                                3
                                   12
                                                     3
                                                        12
               Measuring temperature range: 32°F-100°F (0°C-37.8°C)
               Measuring humidity range: 10% -90% RH (above 65% RH will affect the instrument measuring accuracy)
               Display mode:LCD display.
               Two-stage test voltage: 10V (the measured resistance is less than 10 )/100V (the measured resistance is >10 )
                                                              6
                                                                                         6
               AC adapter: input voltage 120V/AC or 220V/AC, output voltage 12V/DC.
               Power supply: DC or battery power supply, both are common used in US and overseas countries (Note: External power supply and
               battery power supply cannot be used at the same time).
               2.27 kg±.06 kg (5 lbs±2 ounces) cylindrical electrodes with two conductive contact surfaces with a diameter of 63.5 mm±.25 mm (2.5
               inches±.1 inches),The Shore A (IRHD) hardness of the conductive material of each electrode is between 50 and 70.
               Dimension: 186mmx98mmx28mm
               Automatic power-off function to extend battery using life.
               The function of automatic return to zero which is convenient for continuous and quick testing.





                                                     Measurement accuracy
                                                     10E+3 ~ 10E+8: +/-20%@RH<90%
                                                     10E+8 ~ 10E+9: +/-25%@RH<90%
                                                     10E+9 ~ 10E+10: +/-25%@RH<60%
                                                     10E+10 ~ 10E+11: +/-25%@RH<50%
                                                     10E+11 ~ 10E+12: +/-25% or above@RH<40%




               001-9802


               Description
               Measuring resistance: 10  -10 Ω / Surface resistivity: 10  -10 Ω/m 2
                                                     3
                                   12
                                3
                                                        12
               Measuring temperature range: 32°F-100°F (0°C-37.8°C)
               Measuring humidity range: 10% -90% RH (above 65% RH will affect the instrument measuring accuracy)
               Display mode:LCD display
               Two-stage test voltage: 10V (the measured resistance is less than 10 )/100V (the measured resistance is >10 )
                                                                                         6
                                                              6
               AC adapter: input voltage -120V/AC or 220V/AC, output voltage -9V/DC
               Power supply: DC or battery power supply, both are common used in US and overseas countries (Note: External power supply and
               battery power supply cannot be used at the same time)
               2.27 kg±.06 kg (5 lbs±2 ounces) cylindrical electrodes with two conductive contact surfaces with a diameter of 63.5 mm±.25 mm (2.5
               inches±.1 inches),The Shore A (IRHD) hardness of the conductive material of each electrode is between 50 and 70
               Dimension: 191mmx102mmx39mm
               Automatic power-off function to extend battery using life.
               The function of automatic return to zero which is convenient for continuous and quick testing.




                                                     Measurement accuracy
                                                     10V accuracy            100V accuracy
                                                     10E+3 ~ 10E+4: +/-9%@RH<90%  10E+6 ~ 10E+8: +/-5%@RH<90%
                                                     10E+4 ~ 10E+6: +/-5%@RH<90%  10E+9 ~ 10E+10: +/-9%@RH<60%
                                                                             10E+10 ~ 10E+11: +/-15%@RH<50%
                                                                             10E+11 ~ 10E+12: +/-25%@RH<40%